Logo image
Screening for Auditory Dysfunction in Infants by Evoked Oto-Acoustic Emissions
Article de revue scientifique

Screening for Auditory Dysfunction in Infants by Evoked Oto-Acoustic Emissions

Pierre Bonfils, Alain Uziel et Rémy Pujol
Archives of otolaryngology--head & neck surgery, Vol.114(8), p.887-890
01/08/1988
PMID: 3390333

Résumé

• Auditory threshold using auditory brain-stem responses (ABR) was determined in 30 ears from normally-hearing infants and 16 ears from infants with sensorineural deafness. In the same population, evoked oto-acoustic emissions (EOAEs) in response to a click of 20-dB hearing level were recorded. The presence of EOAEs was correlated with ABR thresholds. Evoked oto-acoustic emissions were always present when ABR wave V threshold was equal to or below 30-dB hearing level. On the contrary, infants with ABR thresholds higher than 40-dB hearing level never had EOAEs. As the recordings of EOAEs could be obtained more rapidly than ABR thresholds (average duration: five minutes vs 40 minutes), EOAEs could hold some promise as an objective, easy, and noninvasive test for screening auditory dysfunction in infants.(Arch Otolaryngol Head Neck Surg 1988;114:887-890)