Does repeated disinfection of the e-Flow rapid® nebulizer affect in vitro performance?
Article de revue   Avec comité de lecture

Does repeated disinfection of the e-Flow rapid® nebulizer affect in vitro performance?

Lasha Bakuridze, Véronique Andrieu, Christian Dupont et Jean-Christophe Dubus
Journal of cystic fibrosis, Vol.6(4), pp.309-310
01/07/2007
PMID: 17276148

Résumé

Cystic fibrosis Disinfection Inhaled antibiotics Nebulizer
Repeated disinfection of a nebulizer may modify its performance. During 60 cycles of disinfection with the Nuk® steam sterilizer (120 nebulizations), the median volume diameter of tobramycin or colistin, nebulization time and various parameters correlated with the membrane weight of the e-Flow rapid® are not modified. Results are similar when tap or purified water is used for washing and disinfection.