Résumé
A national retrospective survey on patient doses was performed by the French Society of Medical physicists to assess reference levels (RLs) in interventional radiology as required by the European Directive 2013/59/Euratom.
Fifteen interventional procedures in neuroradiology, vascular radiology and osteoarticular procedures were analysed. Kerma-area-product (KAP), fluoroscopy time (FT), reference air kerma and number of images were recorded for 10–30 patients per procedure. RLs were calculated as the 3rd quartiles of the distributions.
Results on 4500 procedures from 36 departments confirmed the large variability in patient dose for the same procedure. RLs were proposed for the four dosimetric estimators and the fifteen procedures. RLs in terms of KAP and FT were 90 Gy cm2 and 11 min for cerebral angiography, 35 Gy cm2 and 16 min for biliary drainage, 70 Gy cm2 and 6 min for lower limbs arteriography and 70 Gy cm2 and 11 min for vertebroplasty. For these four procedures, RLs were defined according to the complexity of the procedure. For all the procedures, the results were lower than most of those already published.
This study reports RLs in interventional radiology, based on a national survey. Continual evolution of practices and technologies requires regular updates of RLs.